infusion pump equipment composition: syringe pump main unit, power cord
Infusion pump operating conditions: +5°C-+40°C
Operating Relative Humidity: 20%-90%
Atmospheric pressure: 860hPa-1060hPa
Dimensions: 345mm×160mm×135mm (L×W×H)
Supply voltage AC: 220V± 22V/50Hz±1 Hz
Battery type: DC6V rechargeable NiMH battery, working time: ≥3 hours(5ml/h)
Noise: The sound is higher than 65 decibels when the alarm is alarmed, and the sound is lower than 35 decibels when the infusion pump is running normally
Infusion pump alarm method: sound, light, text
Alarm Type:
■ Infusion drug will be exhausted alarm (the remaining amount is less than 10% of the total amount of syringe or the remaining amount of injection time is less than 5 minutes)
■Chemical residue alarm
■Incorrect clamping of syringe and incorrect clamping of push rod alarm
■Power on self-test and forgetting operation waiting for timeout alarm
■Syringe pushing/blocking, reaching the limit alarm, blocking pressure drop alarm
■Alarm for battery undervoltage and network power interruption
■Alarm after the drug infusion is completed