1. the choice of infusion pump tubing

1. It is recommended to use consumables matching the infusion pump, and the infusion pump tube should be used as required and replaced daily.


2. Select the infusion pump tube according to the nature of the infusion liquid and drug, such as the infusion pump tube protected from light.


3. When the infusion rate < 0.5ml/h, a special infusion pump tube with anti-backflow function must be selected to prevent blood return from blocking the pipeline.

Microcomputer Infusion Pump

2. Installation method of infusion pump tubing

1. Insert the infusion pump tubing into the groove on the side of the handle panel.


2. Press down on the handle of the clamp, load the pump tube into the tank, and then loosen it.


3. Install the pump tube into the groove in the middle of the air sensor through the peristaltic extrusion groove.


4. Close the pump door and open the pump tube flow adjustment clamp.


3. Methods for changing the infusion rate

1. Press the [Start/Stop] button to display the stop


2. Press the clear/confirm button, and when the horizontal line flashes below the infusion speed, press up and down to adjust the infusion speed.


3. Press the [Start/Stop] button to restart the infusion, indicating that the infusion is in progress. Note: Decimal points and units


4. Cleaning and disinfection

1. Wipe the shell of the infusion pump with a slightly damp clean soft cloth.


2. The drip rate sensor is cleaned with absolute ethanol.


3. Disinfect regularly to prevent cross-infection, and wipe the instrument with disinfectant such as 5% chlorboxide, 2.25% glutaraldehyde or 75% alcohol.


5. Daily management


1. Conduct a start-up inspection of the infusion pump once a week to test the performance, flow, capacity and blockage pressure of the infusion pump.


2. Avoid liquid seeping into the pump; When the infusion pump is not in use, store it in a cool, dry place away from severe vibration, direct sunlight or ultraviolet radiation.


3. Dedicated management, the establishment of use registration, regular inspection, maintenance and repair system.


Sixth, maintenance

1. Before first use, or after not using it for a long time, when using it again, connect the infusion pump with AC power to charge the built-in battery for at least 12 hours.


2. When not in use for a long time, the internal battery should be charged and discharged at least once a month; The working time is shortened after the battery is charged, and the new battery should be replaced in time; If the infusion pump fails, it should be repaired in time.


Disclaimer: